BKQ-B75II | Vertical Pressure Steam Sterilizer

Product Introduction:

Vertical sterilizer, widely used in laboratories, inspection departments, food processing enterprises, pharmaceutical factories, etc. at home and abroad, for sterilizing medical waste such as instruments, culture media, etc. Vertical pressure steam sterilizer uses gravity displacement to remove cold air from the sterilization chamber, ensuring sterilization effectiveness.

Technical features:

1. Rated working pressure 0.23Mpa,

2. Rated working temperature 134 ℃

3. Operating temperature: 105-136 ℃

4. The product complies withYY/T 1007-2010 standard, and testing reports recognized by the National Food and Drug Administration can be provided.

5. The sterilization chamber and sterilization basket are both made of high-quality stainless steel SUS304 material, with internal polishing treatment and internal circulation of soda water.

6. The handwheel type sliding door structure is equipped with a door safety interlock device and a door detection device. When there is pressure, the door cannot be opened, and the program cannot run if the door is not closed properly.

7. It has anti dry burning alarm, overpressure self release, overtemperature protection, and power safety protection. All alarms have sound and light warnings.

8.LED digital display of sterilization chamber temperature, time, and fault alarm codes.

9. Self expanding silicone rubber sealing ring, with good sealing effect and long service life.

10.The solenoid valve uses internationally renowned brands, and pressure gauges and safety valves are provided with mandatory information such as numbers, nameplates, and certificates of conformity in accordance with national standards.

11.Microcomputer control, with five fixed programs for instruments, dressings, liquids, two custom programs, and a drying function.

12.The entire process of equipment water injection, temperature rise, sterilization, exhaust, and drying is fully automated, with sound and light reminders after sterilization is completed.

13.Temperature uniformity of sterilization chamber: ≤ 2 ℃, drying temperature range: 50~120 ℃.

14. Pulsating exhaust technology ensures steam saturation.

15. Fully protected door cover, with hinges and shafts not exposed.

16. It has fast and slow exhaust functions to avoid the overflow of sterilization liquid.

17. Equipped with a quick repair window, electrical maintenance does not require dismantling the outer cover.


model

BKQ-B50Ⅱ

BKQ-B75Ⅱ

BKQ-B100Ⅱ

BKQ-B120Ⅱ

volume

50L

75L

100L

120L

power

4400W

4700w

5200w

5200W

Net weight

80kg

85kg

100kg

120kg

Internal cavity size

F386x515mm

F386x69mm

F386x875mm

F386x1055mm

External dimensions

550x640x980mm

550x640x1080mm

550x640x1280mm

550x640x1380mm

Number of baskets carried

1

2

2

3

Basket size

F365×360

F365×265

F365×360

Φ 365 × 360 (1 piece)

Φ 365 × 265 (2 pieces)