FH1000PD | Desktop ductless PP fume hood

1、 Technical parameters

1. Basic parameters of fume hood:

(1) External dimensions ≥ (L × D × H) 1000mm × 700mm × 1100mm;

(2) The internal dimensions are ≥ (L × D × H) 980mm × 606mm × 688mm.

(3) Inlet wind speed: 0.3-0.8m/s;

(4) System exhaust air volume: 220~230 m ³/h;

(5) Rated power: 150 W;

(6) Noise level: ≤ 65dB (A);

(7) Illuminance: ≥ 400lx;

(8) Front window glass opening height: 640mm;

(9) LED fluorescent lamp power: 15W;

2、 Product concept:

In the chemical laboratory, various harmful odorous gases, odors, moisture, and corrosive substances are generated during experimental operations. In order to protect the safety of users and prevent the spread of pollutants from the experiment to the laboratory, a fume hood should be used near the pollution source.

The FH700 (PD) desktop ductless PP fume hood, with its excellent acid and alkali resistance, can be used for higher strength acid and alkali experiments without any external pipelines. It can be directly placed on the laboratory desktop for use, making it more convenient for customers to transport, install, and use.

3、 Structural and functional characteristics

1. The outer shell of the fume hood is made of 8mm PP board, which has good smoothness and corrosion resistance.

2. The operation area panel of the fume hood adopts 12.7mm solid psychological board, which can be disassembled for easy cleaning.

3. The front window of the fume hood has an 8 ° physiological slope, which is designed ergonomically and has a larger viewing angle. It is more convenient for operators to use and has a higher aesthetic appeal.

4. The front window of the fume hood is a foldable acrylic window, and the lower acrylic window has the function of stopping at will.

5. The control panel of the fume hood adopts a touch sensitive switch, which integrates the power button, fan button, fan speed control button, and fluorescent lamp button of the fume hood. It is equipped with an LED three digit digital display screen to display the fan gear, making the machine look beautiful and easy to operate.

6. The electrical control system of the fume hood has functions such as overload prevention and electric shock prevention, with stable performance and long service life.

7. The fume hood is equipped with an AC backward centrifugal fan, which has a large air volume and low noise.

8. The fume hood has a power-off memory function, which means that in the event of a sudden power outage, it can be powered on again to maintain the working state before the power outage, facilitating experimental operations.

9. Optional 15W three proof lamp, with no dead corners in lighting, waterproof, dustproof, explosion-proof, no flicker, safer eye protection.