BSC-1500IIA2-S | Double leaf Biosafety Cabinet

Product Introduction

The double door design provides two front openings for two people to operate facing each other at the same time. The front window glass adopts double-layer laminated explosion-proof safety glass. Even if the glass is damaged, it will not harm people, and the biological safety cabinet can still work normally until the end of the experiment, which better protects the safety of personnel and the experiment.

application area

It can be applied to both conventional safety cabinet scenarios and animal research scenarios, where a large number of animals and materials pass through the barrier area without the need to install transfer windows, improving the efficiency of personnel material transfer.

technical parameter

Basic parameters of safety cabinet:

1. Classification: A2 type, 30% external discharge, 70% circulation

2. External dimensions ≥ (L × D × H): 1500 * 785 * 2220mm;

3. Internal dimensions ≥ (L × D × H): 1220 * 700 * 660mm;

4. Height of countertop from ground: 750mm (size can be customized and modified according to requirements)

5. Wind speed: Average descending wind speed: 0.33 ± 0.025m/s; The average suction inlet wind speed is 0.48 ± 0.025m/s;

6. Total exhaust volume of the system: 840 m ³/h

7. Rated power: 1800W (including 500W load on the operating area socket)

8. Noise ≤ 67dB (A)

9. Illuminance (Lx): ≥ 1000lux;

10. Cleanliness level: ISO 4 (Class 10);

11. Filtration efficiency: Both the air supply and exhaust filters use ULPA high-efficiency filters made of world-renowned brand borosilicate glass fiber material, with a filtration efficiency of ≥ 99.9995% for 0.12 μ m particles

12. Number of users: 1-2 people

Structural and functional characteristics:

1. There are two front openings for easy operation, which can be operated by two people facing each other at the same time. A laminar flow air curtain is installed at the front opening to protect the operator and the sample.

2. It has a utility model patent for installing and fixing high-efficiency filters, which facilitates the installation and replacement of high-efficiency filters.

3. The cabinet adopts a tilted angle design, which conforms to ergonomic principles, has a larger viewing angle, is easy to operate, and is more user-friendly.

4. The side wall panels of the exposed work area of the safety cabinet are made of high-quality 304 # stainless steel integrated structure, and the internal washable parts are treated with 8mm rounded corners, leaving no dead corners and easy to clean.

5. The workspace adopts a negative pressure surrounding design, providing better protection and greater safety.

6. The workbench is made of high-quality 304 # stainless steel and adopts a basin shaped design. Even if there is waste liquid overflowing from the experiment, it will not flow into the liquid accumulation tank, making it easy to clean.

7. Fuma caster design: The caster and bracket are integrated, and the safety cabinet can be safely moved through the caster, or fixed and leveled by adjusting the caster support feet.

8. The cabinet and bracket can be separated, and the height of the bracket can be customized and modified according to the actual situation.

9. Reasonable structural design: The maintenance and replacement of the safety cabinet filter and fan can be carried out on the front side of the safety cabinet, which is more convenient and efficient.

10. The glass adopts double-layer laminated explosion-proof safety glass; Even if the glass is damaged, it will not harm people, and the biosafety cabinet can still work normally until the end of the experiment, better protecting the safety of personnel and the experiment.

11. 7-inch touch screen, displaying dynamic airflow direction, displaying descending airflow velocity and inflow airflow velocity in the operation area, displaying temperature and humidity in the operation area, displaying resistance of air supply and exhaust filters, and clock display; All operational status is displayed at a glance.

12. Power outage recovery function: key memory function, fan, sunlight, lighting with power-off memory function. If the power is turned on again after power outage, the corresponding function key will automatically skip the password and enter the main interface, and there will be a power-off alarm prompt.

13. Lock screen function: Touch the lock screen button to lock the operation interface, and other function keys will be locked at this time; Prevent misoperation.

14. It has the functions of timed switch and timed shutdown for fans, ultraviolet radiation, and sockets; Saved work time and improved work efficiency.

15. User password modification: Users can set the startup password for the secure cabinet.

16. By selecting excellent fans and high-efficiency filters, and designing a reasonable air duct structure, the airflow velocity of the biological safety cabinet needs to meet the requirements, and the airflow also needs to achieve directionality, stability, and uniformity to meet the requirements of biological hazard testing.