BSC-1300IIA2-L | Laboratory Biosafety Cabinet

1. Scope of application

Class II A2 safety cabinet is a safety cabinet with a front window operation port. The operator can operate inside the safety cabinet through the front window operation port to protect personnel, products, and the environment during the operation process.

Work environment:

(1) Only applicable indoors;

(2) Environmental temperature: 15 ℃~35 ℃;

(3) Relative humidity: ≤ 75%;

(4) Atmospheric pressure range: 70kPa to 106kPa;

(5) Power supply: 220V 50Hz.

2. Technical parameters

型号 BSC-1300ⅡA2-L 

Power requirement: 220V~50Hz

Dimensions 1300 × 750 × 2250 (mm)

Operation area size 1150 × 600 × 660 (mm)

Rated power 1300W

The total exhaust volume of the system is 440m3/h

UV lamp power 30W

LED light power 12W * 2

Descending airflow 0.33m/s

Inflow airflow of 0.53m/s

Filtration efficiency 99.9995% (for particles with a diameter of 0.12 μ m)

Noise ≤ 67dB (A)

Note: (1) The power consumption of the power supply includes the power of the operating area load (the load cannot exceed 500W);