1. Scope of application
Class II B2 safety cabinet is a safety cabinet with a front window operation port. The operator can operate inside the safety cabinet through the front window operation port to protect personnel, products, and the environment during the operation process.
Work environment:
(1) Only applicable indoors;
(2) Environmental temperature: 15 ℃~35 ℃;
(3) Relative humidity: ≤ 75%;
(4) Atmospheric pressure range: 75 kPa to 106 kPa;
(5) Power supply: 220V 50Hz.
2. Technical parameters
型号 BSC-1500ⅡB2-L
Power requirement: 220V~50Hz
Dimensions 1500 × 760 × 2250 (mm)
Operation area size 1350 × 600 × 660 (mm)
Rated power 1800W
The total exhaust volume of the system is 1270m3/h
UV lamp power 40W
LED light power 16W * 2
Descending airflow 0.33m/s
Inflow airflow of 0.53m/s
Filtration efficiency 99.9995% (for particles with a diameter of 0.12 μ m)
Noise ≤ 67dB (A)
Note: (1) The power consumption of the power supply includes the power of the operating area load (the load cannot exceed 500W);