11231 BBC 86 Pro | Laboratory Biosafety Cabinet

1. Scope of application

Class II A2 safety cabinet is a safety cabinet with a front window operation port. The operator can operate inside the safety cabinet through the front window operation port to protect personnel, products, and the environment during the operation process.

Work environment:

(1) Only applicable indoors;

(2) Environmental temperature: 15 ℃~35 ℃;

(3) Relative humidity: ≤ 75%;

(4) Atmospheric pressure range: 70kPa to 106kPa;

(5) Power supply: 220V 50Hz.

Note: This product is only for use in disease control centers or laboratory research and is not intended for use in clinical in vitro testing laboratories.

2. Technical parameters

This product belongs to Class II A2 biosafety cabinet, which fully complies with the requirements of the Chinese pharmaceutical industry standard YY 0569-2011 "Class II Biosafety Cabinet". The basic technical parameters are as follows:

Model 11231 BBC 86 pro

Rated voltage 220V

Rated frequency 50Hz

Dimensions 700 × 660 × 1925mm

Operation area size 600 × 500 × 540 mm

Rated power 1000W

The total exhaust volume of the system is 230m3/h

UV lamp power 15W

Fluorescent lamp power 8W * 2 (LED)

Descending airflow 0.33m/s

Inflow airflow of 0.53m/s

Filtration efficiency 99.999% (for particles with a diameter of 0.3 μ m)

Noise ≤ 67dB (A)

Note: (1) The power consumption of the power supply includes the power of the operating area load (the load cannot exceed 500W);